Shimadzu Gc-2010 Service Manual
If the flame fails to ignite or the baseline is noisy, remove the FID collector assembly. Use the specialized jet removal tool to unscrew the jet. Inspect it under magnification for cracks or blockages. Clean it ultrasonically in hexane or replace it.
: Inspect for dirt and silica wool position before every analysis series. Cleaning or replacement must only occur when temperatures are below 50∘C50 raised to the composed with power C to prevent burns. shimadzu gc-2010 service manual
Keep in mind that working on complex analytical instruments like the GC-2010 requires specialized training and expertise. If you're unsure about performing repairs or maintenance, it's best to consult a qualified service technician or Shimadzu support. If the flame fails to ignite or the
Platinum resistance temperature detector (PRT/PT100) failure. Check resistance of the oven temp sensor (should be ~100 Ωcap omega at 0°C). Replace sensor if open-circuit or shorted. Clean it ultrasonically in hexane or replace it
To tailor this information to your specific system or issue, let me know:
Caused by dirty carrier gas lines or septa bleed.